UPS Maintenance Bypass Panels isolate the Uninterruptible Power Supply during routine maintenance and emergency servicing while maintaining power to the imaging system. This function also enables the system to remain in use should there be a UPS failure.
Our UPS bypass panels provide the ability to safely lock out power to the UPS. In normal operation, the UPS is online and providing power to diagnostic imaging equipment; in maintenance bypass mode, the UPS is inoperative and can be completely isolated and safely shut down for maintenance.
Bevco UPS maintenance bypass panels are useful for all diagnostic and surgical imaging modalities: X-Ray, MRI, CT, and PET-CT. They’re compatible with equipment from all manufacturers: simply tell us the make and model of your system and we’ll provide the right bypass panel for your needs.
We can also design and manufacture custom UPS bypass panels to meet your unique specifications. Custom configurations can include multiple load breakers to feed multiple imaging systems.

Features & Benefits:
- UPS input and output breakers provide branch circuit overcurrent protection
- Meets disconnection, isolation, and OSHA Lock Out/Tag Out (LOTO) provisions
- 25kA short-circuit current rating or higher when specified
- UL and cUL listed, labeled, and tested; conforms to National Electric Code
- Includes neutral and ground bars
- Panels provide contacts to signal the UPS to switch to static bypass mode
- Circuit breakers are fitted with Kirk-Key interlocks to prevent catastrophic damage to imaging systems and UPS equipment
- Reduce installation time by providing a pre-designed and tested system, eliminating the need to mount and wire numerous individual components
- Factory testing ensures high product quality and reliability
- Installation and operations manuals are included
- Made in the USA
Configuration:
Ordering Information:
Basic Part Number |
kVA Rating |
Amperage @ 480V3PH |
Nominal Dimensions |
||
Height |
Width |
Depth |
|||
UB100-120 |
100 kVA |
120 |
24" |
15" |
9" |
UB150-180 |
150 kVA |
180 |
24" |
15" |
9" |
UP225-270 |
225 kVA |
270 |
24" |
15" |
9" |
UB300-360 |
300 kVA |
360 |
30" |
24" |
9" |
UB100-600 |
500 kVA |
600 |
30" |
24" |
9" |
Listed here are common options and configurations. A wide variety of potential combinations of voltage, short-circuit current ratings, and amperage requirements can be accommodated. Please contact us with your specific needs.
See In-Use Warning Light Controllers with Room Lighting Control Option