CT Main Disconnect Panel provides emergency shutdown, overcurrent protection, OSHA Lock Out/Tag Out provision in an OSHPD Pre-approved stand-alone cabinet. The MDP also serves as a local disconnect for the Diagnostic Imaging System.  Optionally, the Main Disconnect Panel allows for an automatic reclosure after a power outage so the system can come back online without maintenance intervention.

Available in multiple configurations to meet the needs of any CT suite.

  • Emergency Power Off function

Fail-Safe Under-Voltage Trip
~ Opens the Main Circuit Breaker when the EPO is pressed
~ Also provides low voltage protection from power sags
~ Circuit Breaker needs to be reset after EPO event or power sag

Traditional Shunt Trip
~ Opens the Main Circuit Breaker when the EPO is pressed
~ Power to operate the shunt trip is derived inside the MDP
~ Circuit Breaker needs to be reset after EPO event

Contactor-Based NO-TRIP
~ Opens a contactor to remove power from the system when an EPO is pressed
~ Power will be restored when releasing the EPO. No Circuit Breaker needs to be reset

  • Optional integration of power for the chiller into MDP
  • Wide range of voltage and power options to support the smallest CT system to large Dual Generator systems
  • Integral X-Ray ON, X-Ray Ready and Power ON light control available in the MDP
Bevco Main Disconnect Panel for CT

Advantages:

  • Reduce installation time by providing a single-point power connection
  • Proven components and designs eliminate guesswork
  • Standard configurations suitable for use in 25ka short-circuit current rated (SCCR) power systems
  • Standardized design and testing ensures high product quality and system reliability
  • Includes installation and operation manual
  • UL and cUL listed, labeled, and tested; conforms to National Electric Code
  • Surface or semi-flush mounting
  • High SCCR ratings and a variety of custom configurations are available
    Please contact Bevco Imaging Suite Controls for consultation.
  • Seismic ICC-ES-AC156 shake tested approval per OSHPD requirements
  • Panel disconnect provides OSHA lock-out / tag-out provisions
  • Supplied with remote red Emergency OFF guarded Push-to-Latch/Twist-to-Release push button, wall plate and nameplate
  • Factory wired and tested; all devices are selected for high reliability and long life
  • Door has provisions for padlocking to control access
  • Available International CE, 380/400/415V designs include main circuit breaker 30ma, adjustable, RCD protection

 

Configuration:

ct main disconnect panel config diagram

Ordering Information:

Basic Panels

Basic

Part Number

3-Phase Voltage

Amperage

Nominal Dimensions

Height

Width

Depth

CT480-090-BB

440/460/480 50/60 Hz

90

24"

15"

9"

CT480-110-BB

440/460/480 50/60 Hz

110

30"

24"

9"

CT480-125-BB

440/460/480 50/60 Hz

125

30"

24"

9"

CT480-150-BB

440/460/480 50/60 Hz

150

30"

24"

9"

CT400-090-BB

380/400/415 50/60 Hz

90

24"

15"

9"

CT400-110-BB

380/400/415 50/60 Hz

110

30"

24"

9"

CT400-125-BB

380/400/415 50/60 Hz

125

30"

24"

9"

CT400-150-BB

380/400/415 50/60 Hz

150

30"

24"

9"

Option

Description

-BB

Basic Main Disconnect Panel

-BU

UPS Interface option.   Fully integrate emergency power off function if an Uninterruptible Power Supply (by others) is included in the system. See below for compatible UPS Bypass Panel for UPS maintenance and safety.

 

System-matched UPS Bypass Panels

Basic Part Number

kVA Rating

Amperage @ 480V3PH

Nominal Dimensions

Height

Width

Depth

UB100-120

100 kVA

120

24"

15"

9"

UB150-180

150 kVA

180

24"

15"

9"

UP225-270

225 kVA

270

24"

15"

9"

UB300-360

300 kVA

360

30"

24"

9"

UB100-600

500 kVA

600

30"

24"

9"

Listed above are common options and configurations.  A wide variety of potential combinations of voltage, short-circuit current ratings, and amperage requirements can be accommodated. Please contact us with your specific needs.

See UPS Maintenance Bypass Panels and In-Use Warning Light Controllers with Room Lighting Control Option